Important features :
|
The SGR 4 pyrgeometer is an instrument for the highest quality scientific measurements. The specially designed meniscus dome provides a 180º field of view with negligible directional response error. A hard-carbon coating on the outside of the dome smoothes the spectral response and provides extra protection to the silicon surface. The excellent thermal stability of the dome construction and coupling to the instrument body eliminates the need for dome temperature measurements or dome shading. SGR4 can be fitted with the CVF4 ventilation unit to further improve its performance.
The SGR 4 has an internal desiccant that will last for at least 10 years. This minimizes maintenance significantly.
The interval for dome cleaning can be extended, and the quality of measurements maximized, by fitting SGR4 with the CVF4 ventilation unit.
The SGR 4 has a RS-485 Modbus® RTU interface, 0 - 1 V or 4 - 20 mA analogue output, improved response time and temperature corrected measurement data. The long-wave net- and downward radiation are directly available over Modbus®. The wide and low power supply range from 5 to 30 VDC makes integration in meteorological and solar energy stations easy. The instrument is extremely robust and comes with 5 years warranty.
Thanks to standardized output and connections of every SGR 4, exchanging instruments for recalibration is easy.
SmartExplorer Windows™ software for data logging, display of data and Modbus® address setting, can be downloaded from the Kipp & Zonen / Ott HydroMet webpage.
Specifications
Spectral range (overall) | 4,4 to 50 µm = 4400 to 50000 nm |
Response time (95%) | < 18 s |
Window heating offset (@ 0 - 1000 W/m²) | < 4 W/m² |
Zero offset B | < 2 W/m² |
Temperature dependence of sensitivity (-10ºC to +50ºC) | < 1 % |
Operational temperature range | -40...+80°C |
Field of view | 180° |
Non-linearity | < 1 % |
Analogue output (-V version) | 0 - 1 V (-250...+250 W/m²) |
Analogue output (-A version) | 4 - 20 mA (-250...+250 W/m²) |
Digital output | 2-wire RS-485 |
Overall accuracy | ± 1 % |