Kritech & Co PGmbH



 
 
 


POM 2

POM-02 has all the features of the POM-01. But with extended waveband ranges into the near infrared. There is growing interest in monitoring aerosols of larger size and this requires measurement at longer wavelengths.

In addition to the optical system with silicon photo-diode detector of the POM-01, the POM-02 sky radiometer has a second optical system with an Indium Gallium Arsenide infrared detector. The filter wheel has 11 wavebands, providing measurements to 2200 nm. It also has additional UV channels.

POM-02 must be connected to a PC running the operating software in order to make measurements and store data and like the POM-01 data can be post-processed using the SKYRAD.pack software from SKYNET. Maintenance is minimal; consisting of regular cleaning of the optical windows and checking the desiccant in the radiometer.

Main features :

  • Direct solar measurements using an integrated solar tracker with active sun sensor correction
  • Diffuse and scattered measurements by programmed sky scans
  • Programmable measurement times & cycles logging of signal through any, or all, filters
  • One optical system for UV and Visible wavelengths, a second for near-infrared measurements
  • Temperature stabilized detectors and filters
Specifications :
Measurement method Filter wheel, photo-diode & InGaAS detectors
Sun tracker tripod base, sun sensor and rain sensor included
Wavelengths 315, 340, 380, 400, 500, 675, 870, 1020, 1627, 2200 nm
Half-power bandwidth 3 nm for 315 nm filter, 10 nm for others nm
Power supply 230 / 115 VAC, optional 24 VDC
Operating temperature range -30 ºC to +35 °C
Low temperature option -50 ºC to +35 °C
High temperature option -30 ºC to +50 ºC

Available documents (PDF files) :
English



 